Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environment.
Emission test
RF emissions
CISPR11
RF emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions IEC
61000-3-3
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such environment.
Immunity test
Electrostatic discharge
(ESD) acc. to
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Comment: U
is the a.c. supply voltage prior to application of the test level.
T
34/40
Compliance
Group 1
The device uses RF energy only for its internal function. Therefore, RF-emission
is very low and it is unlikely that any interference in nearby electronic equipment.
Class B
The device is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Class A
Symmetrical three-phase devices and other devices.
Passed
Guidance and manufacturer declaration - Electromagnetic immunity
IEC 60601 test level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for mains cables
± 1 kV for input and
output lines
± 1 kV voltage
phase – phase,
± 2 kV voltage
phase – earth
< 5% U
, (>95% dip in U
T
for 1/2 period
40% U
, (60% dip in U
T
for 5 periods
70% U
, (30% dip in U
T
for 25 periods
<5% U
, (>95% dip in U
T
for 5 seconds
3 A/m
med 2314
Electromagnetic environment – Guidelines
Compliance level
± 6 kV contact discharge
± 8 kV air discharge
± 2 kV for mains cables
± 1 kV for input and
output lines
± 1 kV voltage
phase – phase
± 2 kV voltage
phase – earth
)
< 5% U
, (>95% dip in U
T
T
for 1/2 period
)
40% U
, (60% dip in U
T
T
for 5 periods
)
70% U
, (30% dip in U
T
T
for 25 periods
)
<5% U
, (>95% dip in U
T
T
for 5 seconds
3 A/m
2015-01-28
Electromagnetic environment –
Guideline
Floors should be wood, concrete or
covered with ceramic tiles. If floors
are covered with synthetic material,
the relative humidity should be at
least 30 %.
The supply voltage quality should be
that of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
)
Mains power quality should be that
T
of a typical commercial or hospital
)
environment. If the user of the device
T
device requires continued operation
)
during power mains interruptions, it
T
is recommended that the device be
)
powered by a UPS (uninterruptible
T
power supply) or a battery.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital environment.