Changing Measurement Units; Other Measurements - Hach senslON+ MM 374 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4. Operation

4.13. Changing measurement units

The factory configuration is as follows:
Channel 1, pH+EC
Channel 2, ISE
To change units or to measure only one parameter with channel
1 follow the scheme:
10-04-2007
17:15
SYSTEM
SYSTEM
Measurement units
Sample ID
Data output
Unit for Channel 2
SYSTEM
SYSTEM
Measurement units:
Measurement Units:
Channel 1: pH + EC
Channel 1: pH + EC
Channel 2: ISE
Channel 2: ISE
Exit
Save Data
SYSTEM
Measurement Units
Parameters
Channel 1:
and units for
Parameter 1: pH
Parameter 2: EC
Channel 1.
SYSTEM
Measurement Units
Channel 1:
Parameter 1: pH
Parameter 2: EC
SYSTEM
Measurement Units
Parameter 2:
EC
NaCl
TDS
SYSTEM
Measurement Units
Parameter 2:
Any unit
SYSTEM
Measurement Units
Channel 1:
Parameter 1: pH
Parameter 2: --
Save changes
etc.
Note: When Channel 1 is configured to measure only pH, the
instrument asks if the temperature from Channel 2 will be used.

4.14. Other measurements

pH measurement
pH can be measured with both Channel 1 and Channel 2, see
"Change of measurement units".
The pH measurement and measure re-programming is carried
out in accordance with the chapters on page 13. Additionally
two measuring limits (superior and inferior) can be programmed.
When the measured value is out of limits, there is an acoustic
warning, in the printed ticket appears "A" beside the measured
value. The calibration is performed according to the chapters on
page 9 and 11.
...by stability
...
Start measurement
Channel 1
Measuring Conditions
SYSTEM
Resolution: 0.01
Measurement Units
Isopotential: pH 7
Channel 2:
Meas.: By stability
pH
Limits
mV
Start measurement
ISE
etc.
Measuring Conditions
Resolution: 0.01
Isopotential: pH 7
Meas.: By stability
Limits
Start measurement
mV measurement
mV can be measured with both Channel 1 and Channel 2, see
"Change of measurement units".
The standard conditions for mV measurements are:
Measuring resolution: 1 mV/0.1 mV
Measuring mode: by stability, in continuous or by time. See
details in "Measure pH+EC", page 12.
Calibration: Several options, the standard is calibration with
redox standard 220 mV (25˚C, 77 ˚F).
To modify the measuring conditions or the calibration follow the
instrument's instruction.
EC (Electrical Conductivity) measurement
Conductivity can be measured only with Channel 1.
The possible units are:
- Electrical conductivity, uS/cm or mS/cm
- Resistivity,
- Salinity, mg/L or g/L
- TDS, mg/L or g/L
The measuring units are selected following the scheme "Change
of measuring units".
The EC measurement and measure re-programming is carried out
in accordance with the chapters on page 13. The calibration is
performed following the chapters on pages 10 and 11. If the units
are Salinity, the programming parameters are identical to EC.
If the units are TDS, additionally a TDS factor should be programmed.
The standard value is 0.64. If required the user can modify it.
Calibration
The calibration is performed with standard solutions independ-
ently on the measuring units EC, salinity or TDS. The calibration
"In a X value", (indirect calibration), is performed in the meas-
uring units.
Programming Limits
Measuring Conditions
Measuring Conditions
Limits:
Limits
Yes
Superior: pH 8.20
No
Inferior: pH 6.05
17