Re-Programming Ph + Ec Measurements - Hach senslON+ MM 374 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4. Operation
4.10. Re-programming pH+EC measurement
Default programming (standard)
Resolution
• 0.01 pH
Isopotential pH
• pHiso 7.00
TC
• linear, TC = 2.00 % / ˚C.
TRef
• 25 ˚C
Measuring mode • By stability.
pH, standard criterion (variation < 0.01 pH in 6 s).
EC, variation < 1 digit in 6 s.
• In continuous.
- The result is stored when the measurement
has finished.
- The result is printed when the measurement
has finished.
10-04-2007
17:15
Important note:
pH+EC MEASURE
- Once the measurement has been re-programmed, it will be performed in accordance with the new parameters.
Channel 1
- It is highly recommended to apply the same stability criteria to measure and calibrate.
...by stability
Measuring Conditions
Resolution: 0.01
...
Isopotential: pH 7
TC: linear, 2.00%/˚C
Start measurement
Channel 1
TRef: 25˚C
Meas.: By stability
Start measurement
Measuring Conditions
Resolution: 0.01
Isopotential: pH 7
TC: linear, 2.00%/˚C
TRef: 25˚C
Meas.: By stability
Measuring Conditions
Resolution: 0.01
Isopotential: pH 7
TC: linear, 2.00%/˚C
TRef: 25˚C
Meas.: By stability
Measuring Conditions
Resolution: 0.01
Isopotential: pH 7
TC: linear, 2.00%/˚C
TRef: 25˚C
Meas.: By stability
Measuring Conditions
Resolution: 0.01
Isopotential: pH 7
TC: linear, 2.00%/˚C
TRef: 25˚C
Meas.: By stability
Measuring Conditions
Start Measurement
Measuring Conditions
Resolution pH:
0.1
0.01
0.001
Measuring Conditions
Isopotencial:
pH 7
Data introduction
Calculate
Measuring Conditions
TC:
Linear
Natural waters
Measuring Conditions
TRef:
20 ˚C
25 ˚C
Other temperature
Measuring Conditions
Meas.:
By stability
In Continuous
By time
Measuring Conditions
Meas.:
By stability
In Continuous
By time
Measuring Conditions
Meas.:
By stability
In Continuous
By time
Start of measurement with new programming.
Other options
0.1 pH / 0.01 pH / 0.001 pH
pHiso 7.00 / Manual introduction / Value calculation.
Linear, TC selectable between 0 and 9.99 %/ ˚C.
Non-linear for natural waters, according to EN 27888
20 ˚C / 25 ˚C / other values
• By stability.
- Rapid criterion (variation < 0.02 pH in 6 s).
- Standard criterion (variation < 0.01 pH in 6 s).
- Strict criterion (variation < 0.002 pH in 6 s).
• In continuous. Selection of the time interval for
storage or data printing.
• By time. Selection of the time interval for storage or
data printing.
See Isopotential, page 21.
Measuring Conditions
TC linear:
2.00 %/˚C
Save data
Measuring Conditions
Measuring Conditions
By Stability
By Stability
Criteria
Criteria
Rapid
Rapid
Standard
Standard
Strict
Strict
Measuring Conditions
Measuring Conditions
In Continuous
In Continuous
Acquis. Interval
Acquis. Interval
01:00
01:30
Measuring Conditions
Measuring Conditions
By time
By time
Measurement time
Acquis. Interval
00:01:00
00:01:00
Data storage in
the Data Logger. It
coincides with the
printing and PC
sending. Select-
ing 0 store/print
at the end of the
measurement.
13

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido