New tubing, valves and other sample conditioning equipment can be contaminated with oils or dust.
These may contribute to slightly high readings until they are cleaned.
1. Flush the sample line with sample for one to two hours.

Clean the sample components

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Put the analyzer in shutdown mode. Refer to
Clean the sample cells, the stir bar and the sample holder as necessary. Refer to
schedule
on page 4. Refer to the illustrated steps in
and to
Clean the sample holder
8 English
C A U T I O N
Put the analyzer in shutdown mode
Clean the stir bar and sample cells
on page 10.
on page 7.
Maintenance
on page 9