CAUTION!
Prerequisites for operation
Possible consequence: Slight or
minor injuries, material damage
–
The sensor may only be used in
flow gauges that ensure the cor‐
rect flow parameters.
–
There must be a free outlet or at
most a counter pressure of 1 bar
at the outlet of the flow gauge.
The maximum operating pressure
of the respective single compo‐
nents must be observed.
–
The sensor's voltage supply may
not be interrupted
–
Following longer interruptions to
the voltage supply (> 2 h), allow
the sensor to run-in again and
recalibrate it
1.5
Intended Use
NOTICE!
Intended Use
–
The sensor may only be used to
measure and regulate the con‐
centrations of oxidative, bromine-
containing disinfectant
–
Any other uses or modifications
to the system are prohibited
–
The sensor is not a safety com‐
ponent in the sense of DIN EN
ISO 13849-1:2008-12. If there is
a critical process in your meas‐
urement and control system, then
it is your responsibility to make
sure this process is safe
40