Introduction
CAUTION!
Prerequisites for operation
Possible consequence: Slight or minor inju‐
ries, material damage
The sensor may only be used in flow
gauges that ensure the correct flow
parameters.
There must be a free outlet or at most
a counter pressure of 1 bar at the outlet
of the flow gauge. The maximum oper‐
ating pressure of the respective single
components must be observed.
The sensor's voltage supply may not
be interrupted
Following longer interruptions to the
voltage supply (> 2 h), allow the sensor
to run-in again and recalibrate it

1.4 Correct and Proper Use

NOTICE!
Correct and Proper Use
The sensor may only be used to deter‐
mine and regulate concentrations of
free chlorine
The sensor may not be used in
conjunction with organic chlorine
compounds, such as trichloroisocya‐
nuric acid, or stabilisers, such as
cyanuric acid
Any other uses or modifications to the
system are prohibited
The sensor is not a safety component
12
1.5 Information in the event of an
emergency
n
In the event of an emergency, switch off the
controller
n
If liquid exits from the flow gauge housing
then close the shut-off valves provided by
the customer at their inlets and outlets
n
Observe the safety instructions of the plant
operator before opening the flow gauge
housing
n
In case of proven over-metering, destroy the
excess chlorine using a suitable reducing
agent (e.g. hydrogen peroxide or sodium
sulphite)