Decommissioning and Disposal

User qualification: instructed user,
n
Ä Chapter 1.3 'Users' qualifica‐
see
tions' on page 38
NOTICE!
Decommissioning the sensor
Observe all of the safety information
1.
Disconnect the sensor from the
power source
2.
Depressurise the in-line probe
housing
3.
Loosen the clamping screw
4.
Remove the sensor slowly from the
in-line probe housing
5.
Screw open and empty the mem‐
brane cap over a sink or similar
vessel
6.
Flush the membrane and electrode
with clean water and dry until free
of dust
7.
Loosely screw on the membrane
cap to protect the electrodes
8.
Replace the membrane protection
cap to protect the membrane cap
User qualification: instructed user,
n
Ä Chapter 1.3 'Users' qualifica‐
see
tions' on page 38

Decommissioning and Disposal

NOTICE!
Regulations governing disposal of
used parts
Note the current national regula‐
tions and legal standards which
apply in your country
Dispose of the electrolyte in accordance
with the electrolyte's safety data sheet
ProMinent Dosiertechnik GmbH, Heidel‐
berg will take back decontaminated used
devices providing that they are covered by
adequate postage.