Reference To Standards - Microlife BP W150 IMT Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
Weight:
130 g (including batteries)
Dimensions:
80 x 70 x 70 mm
Measuring procedure: oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V dias-
tolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
2 x 1.5 V Batteries; size AAA
Reference to
EN 1060-1 /-3 /-4; IEC 60601-1;
standards:
IEC 60601-1-2 (EMC)
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
10

Publicidad

Tabla de contenido
loading

Tabla de contenido