Immunity test
IEC 60601 test level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
system is used exceeds the applicable RF compliance level above, the system should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the system.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
LCD PROGRAMMING BASE STATION PATIENT MANUAL
Compliance level
10 V/m (compliance
level adjusted to meet
FDA limits)
26 MHz to 2.5 GHz
Note
At 80 MHz and
800 MHz, the
higher frequency
range applies.
ENGLISH
Electromagnetic environment—
guidance
Recommended separation
distance: d = 0.35
80 MHz to
800MHz
Recommended separation
distance: d = 0.70
800MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range.
Interference may occur in the
vicinity of equipment marked
with the following symbol: