Invacare® HomeFill® II Oxygen System

2.3 Provider Information

Invacare recommends leaving a full cylinder of
oxygen with the patient after setting up the HomeFill
compressor.
Before leaving a HomeFill Compressor with a user, the
following checklist MUST be completed:
q
Make sure the pressure test on cylinder has not expired.
q
Check the EMC section for operational information and
possible RF issues. Refer to EMC Information.
q
Instruct the user on the safe operation of the
concentrator and review all warnings.
q
Instruct the user on the safe operation of the HomeFill
compressor and review all warnings.
q
Leave a full cylinder of oxygen with the user.
q
Leave a copy of all warnings and filling instructions with
the user.

2.4 Double-Insulated Products

CAUTION!
– DO NOT unplug the power cord by pulling on
the cord. To unplug, grasp the plug not the
cord, otherwise damage may occur.
In a double-insulated product, two systems of insulation
are provided instead of grounding. No grounding means is
required on a double-insulated product, nor is a means for
grounding to be added to the product. A double-insulated
product is marked with the words "DOUBLE INSULATION" or
"DOUBLE INSULATED" or the symbol:
12

2.5 EMC Information

WARNING!
Risk of Injury or Damage
To reduce the risk of injury or product damage
from interference with wireless equipment:
– Keep the compressor at least 9.8 ft (3.0 m)
away from wireless communication equipment
such as wireless home network devices, mobile
phones, cordless phones and base stations,
walkie-talkies, etc.
WARNING!
– Use of accessories and cables other than those
specified or provided by Invacare could result
in increased electromagnetic emissions or
decreased electromagnetic immunity of this
equipment and result in improper operation.
Medical Electrical Equipment needs to be installed and used
according to the EMC information in this manual.
This equipment has been tested and found to comply
with EMC limits specified by IEC/EN 60601-1-2 for Type
BF equipment. These limits are determined to provide
reasonable protection against electromagnetic interference
in a typical home healthcare environment. Use of this
equipment in an environment with higher levels of
interference could affect the accuracy of the oxygen
1195663-C