Invacare® HomeFill® II Oxygen System
Compressor cycles on and Off
or
O
Below Normal
2
(yellow) and filling (green) lights are alternating
or
O
below normal (yellow) light goes On and Off
2
O
below normal (YELLOW) light stays On
2
32
Problem
1. Check that the concentrator liter flow setting is at
or below the flow setting listed in 4.5 Turning the
Compressor On and Off, page 24.
If necessary, adjust the liter flow setting (refer to the
concentrator User manual). Wait at least 20 minutes for
the GREEN light to appear. If it does not appear, proceed
to step 2.
2. The concentrator may need service, contact your home
health care provider or Invacare.
1. Ensure the concentrator has warmed up for at least
20 minutes. If it has not been 20 minutes, turn the
compressor Off until the concentrator warms up. If the
YELLOW light still remains on proceed to step 2.
2. Inspect the connection between the compressor and the
concentrator for damage or pinching.
•
If damaged, replace.
•
If pinched, straighten tubing. Turn the compressor
Off (O) for at least 10 seconds. Turn the compressor
On ( I ). If the YELLOW light still remains on after 10
minutes, proceed to step 3.
3. The concentrator may need service, contact your home
health care provider or Invacare.
Solution
1195663-C