Invacare HomeFill II IOH200AW Manual Del Usuario página 38

Ocultar thumbs Ver también para HomeFill II IOH200AW:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 89
Invacare® HomeFill® II Oxygen System
Conducted RF
3 Vrms, 6 Vrms
IEC 61000-4-6
Radiated RF
10 V/m
IEC 61000-4-3
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
38
3 Vrms, 6 Vrms
10 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the
Device including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2√P
d = 0.35√P
d = 0.7√P
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
a
1195663-C

Publicidad

Tabla de contenido
loading

Tabla de contenido