Safety Instructions; Product Description; Cleaning And Disinfecting - Bircher CareMat Traducción Del Manual De Instrucción Original

La alfombrilla sensible a la presión para personas con demencia o riesgo de caída
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 7
CareMat
with 868 MHz EasyWave
Translation of the original instructions
1

Safety instructions

These instructions must be completely read through prior to the commissioning and use of the CareMat. They should be kept safely for future use. Warranty
claims will be invalid in the event of any damage to the product, to goods or persons that is due to failure to observe these instructions.
Instructions for nursing staff
• The CareMat operates purely as a support for the nursing staff,
but is not a substitute for them.
• The functional correctness of the CareMat is to be checked each
time that it is used.

Product description

2
Described radio system
2 1
The CareMat with 868 MHz EasyWave
with the ELDAT RCL07 radio receiver. The RCL07 can receive radio signals
from different ELDAT radio transmitters and relay them to a cable connected
call system. Detailed information on this and further compatible
868 MHz EasyWave
®
appliances is available on the following website:
www.eldat.de
2
5
CareMat A01T E868
CareMat A11T E868
1100 x 700 mm
Housing with radio transmitter
1
Chamfered, wheelchair-accessible edges
2
Non-slip surface with stud texture design
3
Day to day operation
3

Cleaning and disinfecting

4
• The CareMat can be cleaned using commercially available disinfecting
and cleaning agents with max. 70 vol% alcohol.
No corrosive agents should be used.
• The CareMat is neither autoclavable nor can it be disinfected using
steam.
®
transmitter
®
radio transmitter is, compatible
6
4
1
4
1
3
CareMat B01T E868
CareMat B11T E868
700 x 400 mm
Align the CareMat centrally in front
of the bed with the handles under-
neath the edge of the bed so that
missing the mat is impossible.
RCL07 radio receiver connected to
the call system.
The pressure-sensitive bed mat for dementia
patients and persons at risk of falling.
Notes for the installer
• The electrical installation work and commissioning of the product may
only be carried out by a qualified technician. Follow ESD protection rules.
Areas of application and function
2 2
The CareMat was primarily designed for use in retirement and nursing
homes, hospitals and psychiatric wards as well as for sheltered residential
accommodation. The necessary care and supervision of patients with
dementia and of those who are at risk of falling represents a significant
increase in care expenditure for care practitioners. The CareMat is the ide-
al support: Risks of falling and the dangers which arise from wandering or
loss of orientation can be significantly reduced.
The CareMat as a pressure-sensitive sensing device is laid directly in front
of the care bed. The RCL07 is connected to the call system via the plug-in
connection of the call button. As soon as the person leaves the bed and
steps onto the CareMat, this switches on and sends out an alarm signal to
the call system via the RCL07. The wireless signalling system reduces the
hazards due to a tangle of wiring around the care bed.
5
5
3
2
Carrying handle
4
Active switching zone
5
RCL07-radio receiver (optional component)
6
Transport and storage
5
• During transport and storage, any significant bending of the
CareMat should be avoided.
• Store the CareMat in a dry, clean and flat location.
• Do not place any heavy loads on the CareMat and do not stack
more than 10 packed mats on top of each other.
4
1
4
2
3
CareMat C01T E868
4
CareMat C11T E868
1100 x 700 mm,
semi-circular
Should the CareMat not be needed,
it can be pushed underneath the
bed.