10. Insert the mercury bulb ④ (respect the polarity of
the bulb), tighten the locking screws and refit the
lamp holder inside the lamp housing. (Fig. 20)
Do not touch the bulb with bare hands.
11. Connect the cable from the external power sup-
ply to the lamp housing. (Fig. 21)
12. Connect the power cable to the external power
supply. (Fig. 22)
The input voltage for fluorescence
power supply is 110-240Vac.
Please use the standard power ca-
ble provided. Select suitable one
when missing or damaged.
Connect the power supply cor-
rectly, be sure to have a good earth
connection.
Before connecting the power cord,
secure the lamp housing cable to
the power supply.
If the power cord is connected be-
fore, ther may be a risk of electrical
shock.
Disconnect all electrical cables
before installing or replacing the
bulb.
The bulb has an anode and a
cathode of different sizes. Respect
the polarity during assembly.