9.4 Checking the System Joint's Free Movement
Check if the system joint moves freely. If the system joint runs with lateral play, mount the next thicker sliding
washer. If it does not move freely (it is jammed), mount the next thinner sliding washer.
9.5 Mounting the Extension Stop Damper and the Small Cover Plate
1 Turn the system joint upside down and place
the extension stop damper into the bore
(fig. 24).
2 Bring the system joint in extension.
3 Lay the cable for the solenoid (fig. 25).
4 Mount the small cover plate to the back of the
system joint (fig. 26).
9.6 Securing the Screws
The screws are secured after the orthosis has been produced and tried on and before it is handed over to the
patient.
1 Secure the screws for the cover plate with the torque corresponding to the system width and
LOCTITE® 243 medium strength.
2 Let the adhesive harden (final strength after approx. 24 hours).
Position of the Screw
S1 (screw 1, axle screw)
S2 (screw 2)
The screws of the cover plate are not secured with the necessary torque at delivery.
You can also find information on the torque in the openings of the cover plate.
10. Controller
The controller is delivered with the component set and has to be mounted to the orthosis. It receives adjust-
ments from the multi-purpose device/Expert app and commands from the remote control/User app, registers
the patient's movements and controls the NEURO TRONIC system knee joint.
The controller can be used for a unilateral as well as a bilateral treatment. It automatically recognises
whether one or two system knee joints are connected to the controller.
32
fig. 24
fig. 25
System Width
16mm
4Nm
3Nm
fig. 26
20mm
4Nm
4Nm