3. Remove the wipe. Allow the meter surface to dry completely.
4. Discard the used wipes and never reuse them. Wash your hands
thoroughly with soap and water after handling the meter, lancing device
and test strips to avoid contamination.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact customer service at
1-888-307-8188 for assistance.
This device has been validated to withstand 260 cleaning and
disinfection cycles using the recommended disinfecting wipe/towelette.
The tested number of cycles is estimated by 1 cleaning and disinfection
cycle per meter per day for 5 years. the expect life of device. The meter
should be replaced after the validated number of cleaning and
disinfection cycles or the warranty period, whichever comes first.
Stop using the meter if you see any signs of deterioration. For example:
● Meter cannot be turned on
● LCD display cracks or becomes cloudy
● Buttons no longer function
● Meter outer casing cracks
● Data cannot be transmitted
● Color or paint/printing on housing is abnormal
● Scratches or abrasions on meter are higher than acceptable
Please contact customer service at 1-888-307-8188 for a replacement
meter if any of the signs of deterioration are noticed.
32