2.7.2.2 Replacing an Interchangeable Unit
Removing the interchangeable unit is done in reverse order:
Removing the interchangeable unit is only possible as long as the piston is in the lower position
(zero position). Possibly, it may be necessary to press <FILL> first.
Depress the black button on the left, and then pull the interchangeable unit forward (Fig. 18 and Fig. 17).

Programming of the titration unit

2.7.3
The data from the RFID transponder of the interchangeable unit will be read immediately (Fig. 19).
Fig. 19
The input menu for the reagents appears for approx.10 seconds (Fig. 20).The size of the interchangeable unit is
displayed on the left side of the display (here 20 ml). It is recommended to enter here at least the name of the
reagent being used. Confirm «Reagent» with <ENTER>/<OK>.
Fig. 20
Then type the name (possible the concentration) and confirm with <ENTER>/<OK> (Fig. 21).
Fig. 21