GB
Specification
IP2044
Application & mounting
Mini-SQUING will work in most liquids, including coating and aerated liquids and
slurries.
For use in safe area only. Mount in any position in tank or pipe. Mounting is by ¾"
or 1" thread or hygienic fitting.
Construction
Wetside material
Gasket
Dryside materials
Body
LED window
Plug
Plug seal
Ingress protection
rating
Operating conditions
Wetside temp.
Ambient temp.
Wetside pressure
Liquid sg
Liquid viscosity
Switching point
(water)
Hysteresis (water)
Switching delay
Maximum altitude
Maximum humidity
Electrical
Switching mode
Protection
Cable connection
Max. conductor size – 1.5mm
PG9 Cable gland provided – cable dia. 6mm to 8mm.
Earthing
Safety EMC
E.M.C. Directive
L.V. Directive
316L Stainless steel (1.4404)
(1" BSPP model only) Non-asbestos
BS7531 Grade X carbon fibre
with rubber binder
304 Stainless steel with polyester label
Flame retardent Polyamide
(Pa12) UL94 V2
Polyamide glass re-inforced
Nitrile butadiene rubber
IP66 an IP67
o
o
-40
C to +150
C
o
o
-40
C to +80
C
o
(derated to 50
C at 150
-0.25 bar g to +100 bar g at 50
(30 bar for hygienic fittings)
0.6 to 2.0
0.2 to 10,000 cPs
13mm from tip (vertical) / from edge
horizontal) of fork
+/- 1mm nom.
1 sec dry to wet / wet to dry.
2000m
100% RH
User selectable (Dry = on or Wet = on) by
selection wiring in plug
Reverse polarity protected. Missing load /
short circuit protection.
Via 4 way plug provided – DIN43650
Orientation – 4 position (90/180/270/360 deg.)
2
Mini-SQUING must always be earthed
either through cable plug or using
external earth connection provided.
EN50081-1 (Emmissions)
EN50082-2 (Immunity) EN61326
EN61010-1
Pollution degree 2, Category II (264V max)
Pollution degree 2, Category III (150V max)
Page 20
o
C wetside)
o
C