TroubLesHooTIng / speCIfICATIons

TRoUbleshooTInG

Do not attempt to open the device for maintenance or repair. The device contains no user-serviceable parts. Contact your
oxygen provider if service is required. Do not attempt any other maintenance.
Symptom
Oxygen is not being delivered even though the
indicator is flashing every time I breathe.
Use times are different from those stated in
the literature.
PulseDose will not pulse
PulseDose works fine for a couple of minutes,
then sensitivity seems to drift and may stop
working altogether.

sPeCIfICaTIons

iFill Oxygen Cylinder w/Integrated PulseDose Regulator (US) (Canada)
Product # (US)
Product # (Canada)
PD1000A-M4
PD1000A-CAN-M4
PD1000A-M6
PD1000A-CAN-M6
PD1000A-CAN-ML6
PD1000A-ML6
PD1000A-C
PD1000A-CAN-C
PD1000A-D
PD1000A-CAN-D
PD1000A-E
PD1000A-CAN-E
Power Supply ........................................................................................................................................... (2) Standard "AA" alkaline or NiMH batteries.
Power Requirements . ..................................................................................................... Batteries other than alkaline or NiMH are not recommended
Degree of Protection Against Electric Shock . .................................................................................................................................TYPE BF applied part
Modes of Operation . ......................................................................................................................................................................... Continuous / Pulsed
Approval Body And Safety Standard .......................................................................................................... IEC 601-1; CAN/CSA-C22.2 No 601.1-M90
US Patents ..................................................................................................................................................................... 4,519,387; 5,755,224; 4,457,303
� wArnIng
Possible Causes
1. O xygen supply is empty.
1. P ulseDose responds to your breath rate.
Your breath rate may vary, which causes the
operation time to vary.
2. L eak in system
1. C annula is not adjusted properly.
2. U nit is not turned on.
3. B atteries discharged or not installed.
4. M outh breathing with closed soft palates.
5. U nit did not reset while changing batteries
(red light stays on).
1. U sing pediatric cannula or any cannula that
restricts continuous flow capacity of 10 lpm.
Weight
Dimensions
3.5 lbs (1.59 kg)
3.2"D x 12.4"L (81mm x 315mm)
4.1 lbs (1.86 kg)
3.2"D x 15.59"L (81mm x 396mm)
4.8 lbs (2.18 kg)
4.38"D x 11.68"L (111mm x 297mm)
5.6 lbs (2.54 kg)
4.38"D x 14.88"L (111mm x 378mm)
7.2 lbs (3.27 kg)
4.38"D x 20.51"L (111mm x 521mm)
9.8 lbs (4.45 kg)
4.38"D x 29.63"L (111mm x 753mm)
due to the capacity needed for operation and battery life of the unit.
Remedies
1. C heck contents indicator on the device. If
empty, switch cylinders.
1. P ulseDose is operating correctly.
2. C ontact your DeVilbiss provider.
1. C heck all cannula connections to make sure
they are tight, and adjust the cannula to fit
comfortably in your nose. Ensure tubing is
not kinked.
2. T urn the rotary selector to the appropriate
setting.
3. I nstall new batteries.
4. B reathe through the nose (cannula).
5. T urn the unit off and back on using the
rotary selector.
1. R eplace with standard nasal cannula.
A-535-CYL