KEY TO DISPLAY ICONS

DISPLAY MARK
or
CONDITION/CAUSE
Appears when the cuff is
attached too loose.
Appears when the cuff is not
attached correctly and the pulse
cannot be detected.
Appears when the pressure is
300mmHg or higher.
Appears when the batteries have
run out.
Appears when the pulse rate
falls outside the measuring range
(below 39 beats/min. or above
181 beats/min.)
The unit does not function
properly.
- Eng 23 -
CORRECTIVE ACTION
Attach the cuff properly and
measure again.
(
See page 15 "ATTACHING
THE CUFF")
Replace the batteries. (
See page 11 "REMOVING THE
BATTERIES".)
Attach the cuff properly again.
Take deep breaths to relax, and
measure again. (
20 "TIPS FOR OBTAINING
ACCURATE MEASUREMENT".)
Contact the store where it
was purchased or your local
service center. (
26 "BEFORE REQUESTING
REPAIRS OR TESTING".)
See page
See page