KEY TO DISPLAY ICONS

Icon displayed
If any of the following marks is displayed, take the specified action, then
repeat the operation.
Icon displayed
Displayed during
measurement.
Displayed during
measurement.
Condition/Cause
• Appears when the
blood-pressure could
not be measured
accurately.
The pressure is
281 mmHg or higher.
Appears when the
batteries are running low.
Detected problem in the
device.
-
Eng 20
Condition/Cause
Lit
Measurement is under way.
Blinking
Pulse detected.
"PUL Err" is displayed when the pulse
value is below 40 or above 181.
Corrective action
Apply cuff properly and
measure again.
If the pressure does not
automatically decline
during measurement,
immediately press the
"START/STOP" switch
and turn the unit off.
Take the measurement
again.
Replace the batteries.
Contact the vendor.
-