9.12 Centering the condenser
1. Place the specimen on the stage, insert 10x
objective into the light path and focus.
2. Insert the front lens of the swing-out condenser
①. (Fig. 34)
3. Rotate the field diaphragm ring ② clockwise, to
fully close the diaphragm.
4. Rotate the condenser height adjustment knob ③
to focus the edges of the diaphragm.
5. Rotate the two centering screws ④ to bring the
bright spot in the center of the field of view.
6. Gradually open the diaphragm. The condenser is
centered when the diaphragm image is symme-
trical to the field of view.
7. In normal use, open the diaphragm until it cir-
cumscribes the field of view.
9.13 Effect of field diaphragm
Field diaphragm adjusts the illuminated area to obtain
a high contrast image.
Set the diaphragm according to the objective in use
until it circumscribes the field of view, in order to eli-
minate unnecessary light to eyepieces. (Fig. 35)
9.14 Aperture diaphragm
•
The Numerical Aperture (N.A.) value of the aper-
ture diaphragm affects the image contrast. Incre-
asing or reducing this value one can vary resolu-
tion, contrast and depth of focus of the image
•
With low contrast specimens set the numerical
aperture value ⑤ (printed on the condenser ring)
to about 70%-80% of the objective's N.A. (Fig.
36). If necessary, remove one eyepiece and, loo-
king into empty eyepiece sleeve, adjust the con-
denser's ring in order to obtain an image like the
one in Fig. 37.
Example: with objective PLAN 40x / 0.65 set the
scale to 0.65 x 0.8 = 0,52
③
①
②
APERTURE
DIAPHRAGM
FIELD OF VIEW
Page 23
④
Fig. 34
Fig. 35
⑤
Fig. 36
70-80%
30-20%
Fig. 40
Fig. 37