RECETTE
3.7. PERTE A LA TERRE
Mesurer les courants de perte à la terre en toutes conditions, en utilisant l'agencement de circuit figurant ci-
dessous.
Vérifier que les résultats soient conformes au tableau ci-dessous:
S1
S2
off
off
off
on
on
off
on
on
off
off
off
on
on
off
on
on
= VOLTMETRE NUMERIQUE: IMPEDANCE D'ENTREE = 10 MOhm
R = 1 KOhm ± 1%
C = 0.15 µF ± 5%
T = K1
FICHE DU SYSTEME MOBILE
L
=
PHASE
N
=
NEUTRE
P.E. =
TERRE
BARRETTE DE CONNEXION DU SYSTEME MURAL
L
=
PHASE
N
=
NEUTRE
P.E. =
TERRE
D-4
S3
Courant perte max. (mA)
Mobile
on
< 2
on
< 2
on
< 2
on
< 2
off
< 2
off
< 2
off
< 2
off
< 2
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
SYSTEMA ORALIX AC
Mural
< 0.5
< 0.5
< 0.5
< 0.5
< 2
< 2
< 2
< 2
Gendex Dental Systems