SETTING TO WORK
3. CONTROLS, INDICATORS AND JUMPERS SETUP
For a survay of the controls and indicators of the Timers see CZ-1 and CZ-2.
3.1. JUMPERS AND DIP-SWITCHES SETUP
The table below describes the jumpers and dip-switches setup on Densomat/Secondent Timer.
Factory
default setting
W1
Not used
W2
1-2 (B.U.T. enabled)
W3
Unplugged
W4
Plugged
W5
Closed
W6
Opened
W7
Opened
W8
1-2 (X-ray enabled)
W9
Opened
W10
Closed
W11
Closed (F3 shunted) Opened (F3 working)
S1 DENSOMAT
1
not used
2
OFF=
single Tubehead (*)
ON =
double Tubehead
3
ON= compensation
(**) inhibited (*)
OFF = compensation
enabled
4
OFF= X-ray inhibited
ON = X-ray enabled (*) ON = X-ray enabled (*)
8
not used
(*) = DEFAULT (**) = Manual mode only
(Exposure time selected manually by pressing S+/S- pushbuttons)
Preheating time selection at 50Hz
(applicable for Densomat & Secondent)
TIME
S1
(ms)
5
6
7
80
OFF
OFF
OFF
100
OFF
OFF
ON
120 (*)
OFF
ON
OFF
140
OFF
ON
ON
160
ON
OFF
OFF
180
ON
OFF
ON
200
ON
ON
OFF
220
ON
ON
ON
C-2
DENSOMAT
Custom setting
Not used
2 - 3 (B.U.T. test)
Not used
Not used
Not used
Not used
Not used
2-3 (X-ray inhibited)
Not used
Not used
SECONDENT
see table
"Tubehead version
selection"
OFF= compensation
inhibited
ON =compensation
enabled (*)
OFF= X-ray inhibited
not used
Tubehead version selection
(*)= DEFAULT
Tube version 230V (*)
S1 - 1
S1 - 2
OFF
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
Factory
default setting
Not used
1 - 2
Opened
Closed
Closed
Opened
Opened
1-2 (X-ray enabled)
Opened
Closed
Closed (F3 shunted)
(Secondent only )
240V 220V spare
ON
OFF
ON
OFF
ON
ON
OFF
ORALIX AC SYSTEM
SECONDENT
Custom setting
Not used
Not used
Not used
Not used
Not used
Not used
Not used
2-3 (X-ray inhibited)
Not used
Not used
Opened (F3 working)
Gendex Dental Systems