Descargar Imprimir esta página

Mobile System Mounting; Room Preparation; Unpacking - GENDEX Oralix AC Manual De Servicio

Ocultar thumbs Ver también para Oralix AC:

Publicidad

Idiomas disponibles

Idiomas disponibles

ORALIX AC SYSTEM

3. MOBILE SYSTEM MOUNTING

The Mobile
mains

3.1. ROOM PREPARATION

A mating wall socket shall be prepared within 1.5 m from the chair, with the protective earth pin connected to
the equipotential earth point of the room, and with the line wires of the proper size.
The mains resistance, with 230 V A.C. mains, as measured at the wall socket must be
less than 0.618 Ohms.

3.2. UNPACKING

Unpack each box and check the completeness according to ITEMS SUPPLIED BZ-3.
Verify also that the rated voltage of the Timer and Tubehead, as stated on the type number plates, are
corresponding to the local situation.
Gendex Dental Systems
tand incorporate
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
NOTE
INSTALLATION
B-17

Publicidad

loading