PREVENTIVE MAINTENANCE

4.3.2. Inspection of the Timer

(1)
Verify the proper operation of the Timer according to Section C- "Functional Test" of this manual.
(2)
Inspect exposure switch and cable for wear: check for continuity and for intermittent open or short
circuit; replace if necessary.
(3)
Inspect the operator control panel for wear; verify that the LEDS indication over each push button are
working properly.
(4)
Check the adjustment of the mains measuring circuit according to Section C "Ajustment of the mains
voltage measuring circuit" of this manual.

4.3.3. Exposure counter reading

Record on the EZ-1 module the read out value of exposure counter reading.
5. FUNCTIONAL TEST

5.1. PERFORMANCE TEST

See Section C of this manual.

5.2. CUSTOMER TEST

To prove to the customer that the system is in compliance with the standards, perform the tests described in
the Section D of this manual.
Gendex Dental Systems
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
ORALIX AC SYSTEM
E-9