PREVENTIVE MAINTENANCE
4.2.2. Maintenance of the Extension Arm
For the maintenance of the Extension Arm follow the
procedure below:
1
Unscrew the two friction screws (h).
Extract slightly the arm base to access the ball
2
bearing (i): inspect for wear and lubricate it.
3
Screw back the friction screws (h).
4
Remove the snap-out caps (l) and inspect for wear the bush (m): replace it if required and remount the
caps (l).
5
Lift slightly the Extension Arm at the side of the Wall Support and grease the spigot (n).
6
Inspect for wear ring (o) and replace it if required.
7
Inspect for wear the cabling and check the protective earth connection.
4.2.3. Maintenance of the Mobile Stand
For the maintenance of the Mobile Stand follow the
procedure below:
1
Check for wear the wheels (a) and verify the
correct rotation: replace them if necessary.
Inspect the wheel brakes (b) and verify the effi-
2
ciency of their action: replace the complete
wheels with brakes if required.
3
Inspect for wear the mains cable and its attach-
ment to the column.