Menus
Data log
The SG8 is designed to store up to 200 sets of measurement data in the memory.
Auto Save
"Auto Save" logs automatically every endpointed reading
If "Auto Save" is set to Off, the Store softkey appears in the measuring screen. You can then manually store
endpointed data by pressing this softkey.
T-int. read (Timed-interval reading)
With timed-interval reading, a reading is stored to memory every time after a certain interval (3 – 9999 s)
defined in the menu has elapsed. You may stop the reading by pressing Read. When timed-interval reading
is "on", DL is shown on the display.
For readings lasting longer than 15 minutes, switch off the auto-off function by pressing and holding down
for two seconds.
When working in the timed-interval reading mode, you can define the length of the measurement period by
selecting the appropriate endpoint mode ( ,
format".
Data output
Data stored in the instrument's memory can be transferred to the METTLER TOLEDO RS-P42 printer or PC
software (e.g. using BalanceLink) via the infrared interface.
GLP
You can choose between two different data formats when outputting the data from memory: GLP ON or
GLP OFF:
Examples
GLP format
GLP
On
Date
01-JAN-05 / JAN-01-05
Time
09:31:03
Sample_ID
040914
Result
pH 6.986
mV
5.3 mV
Temperature
25.3 °C
ATC/MTC
ATC
Endpoint
Auto
Sensor_ID
04091401
Last_cal.
09-13-04
Signature
________
Non-GLP format
GLP
Off
Result
pH 6.986
Temperature
25.3 °C
ATC/MTC
ATC
Endpoint
Auto
16
,
and
to the memory.
and
) in the menu "2. Set. meas." under the item "1. EP
METTLER TOLEDO SevenGo pro™ pH