5. Setting the treatment values

Warning!
m
Patient hazard –
Before therapy, a test run consisting
of several exercise cycles must be com-
pleted without the patient. Then, repeat
the test run with the patient and check
that the movements do not cause any
pain.
Note!
See also 2.2 and 2.3 as well as page V!
5.1 General information on
programming the
ARTROMOT®-SP3
1. You activate the programming mode by
briefly pressing the MENU button on the
control pendant.
2. The treatment parameters and functions
are allocated to four (Standard model) or
six (Comfort model) programming levels
(four per level).
To be able to program a parameter you
will have to access the corresponding
programming level. This is also done with
the MENU button. You advance one level
each time you press the button. The code
M1, M2, etc. that appears in the middle of
the display indicates the programming
level.
3. You activate the treatment parameters
and functions with the four parameter
buttons below the display. The symbols
above the 4 parameter buttons indicate
the assigned parameters and functions.
This is what happens when you press
one of the parameter buttons to select a
parameter:
– The corresponding symbol appears on
the display in a larger format.
– The set value is displayed.
– The symbol above the parameter but-
ton is highlighted.
4. With the + / - buttons (plus/minus) you
change the displayed value. When you
press and hold the button, the value will
change at a higher rate.
Some of the (special) functions can only
be enabled and disabled. This is done by
pressing the corresponding parameter
button or with the + / - buttons. Active
parameters are identified with a check
mark in the circle next to the symbol.
5. Having programmed all parameters, press
the STOP button to save the values.
6. Next, press the START button: The
ARTROMOT®-SP3 will check the set
values, moves to the middle position and
stops.
7. Press the START button again to start
therapy.
In the synchronized mode, the carriage
will then move from the middle position
to the maximum eversion value. After-
wards, the carriage will simultaneously
move to the maximum inversion value
and to the maximum plantar flexion
value, and then simultaneously to the
maximum dorsal extension value and
the value halfway between inversion and
eversion. After reaching this position, the
cycle restarts: maximum eversion at the
same time as halfway between dorsal
extension / plantar flexion, etc.
After activation of the START button in
the non-synchronized mode, the motors
will perform random movements, each
motor reversing after reaching the maxi-
mum values.
54