LEVEL 3:
– Reverse on load - motor A
– Reverse on load - motor B
– Motor A ON/OFF
– Motor B ON/OFF
LEVEL 4:
– Warm-up protocol
– Isolation protocol
– Dorsal extension /
plantar flexion therapy
documentation
– Inversion / eversion
therapy documentation
LEVEL 5:
– Stretching in dorsal
extension
– Stretching in plantar
flexion
– Stretching in inversion
– Stretching in eversion
LEVEL 6:
– Transport setting
– Synchronized/
non-synchronized mode
− Total therapy time
– Service menu
5.5 Protocol details –
MENU
– To access the appropriate programming
level, press the MENU button repeatedly.
– The treatment parameters are selected
with the different parameter buttons.
– To change the treatment values, press
the + / – buttons.
– To enable/disable a function, press the
MENU
respective parameter button again.
– To save the entered values, press the
STOP button.
All special functions are disabled in the
default settings.
Note!
LEVEL 1: equivalent to level 1 of the
Standard model (see: 5.3)
LEVEL 2: equivalent to level 2 of the
Standard model (see: 5.3)
MENU
LEVEL 3: equivalent to level 3 of the
Standard model (see: 5.3)
LEVEL 6: equivalent to level 4 of the
Standard model (see: 5.3)
LEVEL 4:
n Warm-up protocol
MENU
During warm-up, the patient will slowly
become used to the set maximum dorsal
extension / plantar flexion and inversion /
eversion values, starting from the middle
position.
The warm-up protocol starts in the
middle between the two maximum
values set for dorsal extension / plantar
flexion and inversion / eversion.
In the beginning, the range of inversion
/ eversion motion increases with each
cycle, until the programmed maximum
values are reached after a total of 15
cycles. In this phase, motor A (dorsal
extension / plantar flexion) is inactive.
Then the carriage stops again in the
middle position and starts increasing the
range of dorsal extension / plantar flexion
motion with each cycle, until the pro-
grammed maximum values are reached
after a total of 15 cycles. In this phase,
motor B (inversion / eversion) is inactive.
60
Comfort model