2. Resting the foot by programming a
plantar flexion angle of 0°
3. Programming the maximum ever-
sion value
4. Resting the foot by programming
an inversion angle of 0°
5. Programming the maximum plan-
tar flexion value
6. Programming the maximum inver-
sion value
– When the START button is pressed in
the synchronized mode, the device
will first move from the middle posi-
tion to the maximum eversion value.
Afterwards, the carriage will simul-
taneously move to the maximum
inversion value and to the maximum
plantar flexion value, and then simul-
taneously to the maximum dorsal
extension value and to the value half-
way between inversion and eversion.
After reaching this position, the cycle
restarts: maximum eversion at the
same time as halfway between dorsal
extension / plantar flexion, etc.
LEVEL 2:
n Pauses
Pauses occur at each programmed maxi-
mum value.
These are the three pause points:
maximum dorsal value at the same time
as halfway between inversion / eversion
maximum eversion value at the same
time as halfway between dorsal exten-
sion / plantar flexion
maximum plantar flexion at the same
time as maximum inversion
Pauses are adjustable in steps of 1 second
between 0 and 59 seconds.
Default setting: no pause
n Timer (therapy time)
Default setting is continuous operation.
A clock symbol in the upper right-hand
corner of the display identifies the
continuous mode of operation. The clock
indicates the elapsed therapy time.
In the continuous mode, the device must
be stopped with the STOP button.
However, you can also select therapy
durations of 1 to 59 minutes in steps of
1 minute and of 1 to 24 hours in steps of
30 minutes. When the time has elapsed,
the device switches automatically off
and stops in the middle between the set
limit values.
In this case, a circle replaces the clock
symbol. The circle fills as the therapy time
progresses.
n Speed
The speed can be adjusted in steps of 5 %
between 5 % and 100 %.
Default setting: 50 %
n New patient
With this function, the CPM device will
move to the home position. Select the
function and press the START button.
The device enters the home position
and existing therapy parameters will be
deleted.
On ARTROMOT®-SP3 devices with a
patient chip card the factory defaults
are restored. All values stored on the chip
card will be deleted.
The carriage stops in the home position,
dorsal extension / plantar flexion = 0° and
inversion / eversion = 5°.
The "New Patient" function (home posi-
tion) selects the following settings:
– Dorsal extension:
– Plantar flexion:
– Inversion:
– Eversion:
– Pauses:
– Timer:
– Speed:
– Reverse on load - motor A: 25
– Reverse on load - motor B: 25
– Motor A:
57
20°
- 20°
- 5 °
15°
0
continuous
operation
50 %
ON