5.4 Typical phases of density determinations
Density determinations include the following two phases.
Fill the measuring cell with sample and perform the measurement.
•
Prepare the measuring cell for the next sample.
•
5.4.1 Fill the measuring cell and measure the density
Because the measuring cell always contains residue from the previous sample or the cleaning solution, it is
important that you remove the residue using the new sample. This is only possible if the residue is soluble in the
new sample.
To achieve good results, make sure that the measuring cell contains only the sample you want to measure and
that there are no bubbles in the measuring cell.
You can use the sample pump or syringes to fill, drain and rinse the measuring cell. The sample pump is used
for samples with a viscosity up to 100 mPa*s. Syringes are typically used if the viscosity of the samples is
above 100 mPa*s.
You can find more information on working with the syringe in the Reference Manual.
Handheld Density Meter
1
Max.
25
OK
4
Function
Cycles between the screens with small, capital or special characters.
Deletes all entered characters.
Deletes the character to the left of the cursor.
Moves the cursor.
Displays available characters, numbers and symbols.
Maximum number of characters you can enter.
Minimum number of characters you need to enter.
Displays the entered characters.