Push the button [4] Manual Operation
Select a channel with button [1]
Select a parameter with the central controller [11]
Confirm the selection with confirmation button [10]
Adjust the value with the central controller [11]
Confirm the setting by pushing the confirmation
button [10]
To start the emission push the start button [5] the
Stop symbol ◼ change to running 
To interrupt the treatment push the stop button [6]
once. Running symbol  change to pause
To Stop the treatment push the stop button [6]
again when the pause symbol is active
Push Return button [2] to return to the main menu
NOTE: The Endolaser 422 has two independent output channels. On each channel you can set
parameters. Repeat the operation as described above with the other channel.
CAUTION:
Laser treatments must be delivered under the strict control of the operator to conscious patients
who are able to interact with the operator in response to stresses transmitted by the machine. In
case of non-compliance with the information provided, Enraf-Nonius B.V. will not be held
responsible for any accidents.
It is the responsibility of the operator to verify that the emitting head remains well adherent to the
treatment area, in order to avoid the emission of the laser in areas other than those to be treated.
The laser radiation coming out of the device is dangerous: always use the appropriate protective
glasses. Always avoid eye exposure to direct or diffuse radiation.
It is advisable not to start the treatment delivery if the machine does not show up in perfect
mechanical conditions or if there are no laser emission characteristics that conform to the
purpose (see the table of technical characteristics).
NEVER RETURN THE HANDPIECE TO THE PARTS OF THE SENSITIVE BODY TO THE
LASER RAYS, FOR EXAMPLE OF THE EYES.
DO NOT LOOK AT THE RADIUS ISSUED BY THE HANDPIECE OR THE DIRECT RETURN
WHICH IS REFLECTED DURING THE TREATMENT.
5.7.2

Parameter Adjusting

By rotating the central controller the entire range of alphanumerical characters can be chosen.
The percentage by which the "Pulsed mode" is defined represents the percentage of time the cycle is in
action as opposed to the length of time of the entire working cycle.
Therefore 50% attributes the same time length to the action phase and to the following interval.
Pulsed: 100% is equal to the continuous mode.

5.7.2.1 Select Program via Manual Operation

The memorized programs available have been specially created by professional and expert operators
after years of experience in the field.
A list of all available programs can be found in Appendix B.
When you are in the Manual Operation you have also access to the pre-programmed protocols.
Select a channel with button [1]
Select "Program" with the central controller [11] and accept the selection with confirmation button
[10].
Select from the list the desired protocol with the central controller [11] and confirm the selection
with confirmation button [10]
Start the emission with de start button [5]
EN
12
Select Channel 2
Return
Personalised
Channel < 1 >
Freq.: 5000Hz
Time: 00:20 s
Dosage: 6 J/ cm²

Select Program

LP025
00:00
0,1 J/cm²
12 mWatt
LP100
00:00

Publicidad