CAUTION:
Laser treatments must be delivered under the strict control of the operator to conscious patients
who are able to interact with the operator in response to stresses transmitted by the machine. In
case of non-compliance with the information provided, Enraf-Nonius B.V. will not be held
responsible for any accidents.
It is the responsibility of the operator to verify that the emitting head remains well adherent to the
treatment area, in order to avoid the emission of the laser in areas other than those to be treated.
The laser radiation coming out of the device is dangerous: always use the appropriate protective
glasses. Always avoid eye exposure to direct or diffuse radiation.
It is advisable not to start the treatment delivery if the machine does not show up in perfect
mechanical conditions or if there are no laser emission characteristics that conform to the
purpose (see the table of technical characteristics).
NEVER RETURN THE HANDPIECE TO THE PARTS OF THE SENSITIVE BODY TO THE
LASER RAYS, FOR EXAMPLE OF THE EYES.
DO NOT LOOK AT THE RADIUS ISSUED BY THE HANDPIECE OR THE DIRECT RETURN
WHICH IS REFLECTED DURING THE TREATMENT.
5.7.4

Create programs

Select "Create program" in the main menu by
pressing button [2]
Select a position in the list with the central controller
[11]
Confirm the selection with the confirmation button
[10]
Adjust the parameters
To change the frequency
Select Freq.: with the central controller. Accept the
setting with the confirmation button [10] and adjust
the value with the central controller [11] Confirm with
the confirmation button [10]
Repeat the procedure for:
Pulse mode
Time
Program name
EN
14
Select a position on the line
o
Confirm the position with button [10]
o
Change the characters (number and
o
characters can be changed)
Confirm with button [10]
o
Select another position
o
Press OK button [1] when the name is
o
complete
Return
Select Program
01.............
02.............
03.............
04.............
05.............
06.............
07.............
08.............
09..............
10...............
OK
Create Program
Freq.: 5500Hz
Time: 00:20 s
Dosage: 6 J/ cm²