MEASUREMENTS

(A) Insulation resistance measurement test

1. Connect speciment by test cable..
2. Switch the function rotary switch to turn on power and select test voltage
from 1kV, 2.5kV, 5kV or 10kV respectively.
3. Be sure that the cable connecting the specimen, if there is an exterior
voltage (above AC30V or DC30V) exerted, beeper activates in response;
and LCD displays the warning picture as shown in the AC/DC voltage
measurement function:
Now, test cannot be performed. To go on the test it should remove the exterior
voltage.
4. Then, press TEST / STOP button to test.
(1) Rotate another angle to lock TEST / STOP button for testing
continuously.
(2) While test is running, beeper activates to remind operator that test is
underway.
(3) After the set test time is due (see the part: the test time setting), test
stops and system will automatically lock down the test value.
(4) To read the test value on the LCD display
10kV/20TΩ
5kV/10TΩ
2.5kV/5TΩ
1kV/2TΩ
OFF
V
AC/DC
(30~600V)
- 20 -