Device use

MIO-IONOTENS is a portable and battery-powered device that generates TENS
and ionophoresis currents. It is particularly indicated for daily treatments of
the most commons muscle diseases.
MIO-IONOTENS has 14 preadjusted TENS programs, 10 preadjusted programs
REHA (including 3 programs iontophoresis) and 12 free memories adjustable
by the user to create programs according to his needs. The program MEM 13
is a battery test.

Operating instructions

It is recommended reading the entire user manual before using.
To start the therapy, turn MIO-IONOTENS on using the
PREADJUSTED PROGRAMS
Read the follow instructions to use the preadjusted programs and start the
therapy:
1. Select the program group using the MODE/ESC button (TENS, REHA,
MEM).
2. Select the program using PRG+ and PRG- buttons (refer to the follow
sections to get all technical specifications).
3. Increase current intensity for the channels using CH1 and CH2 (▲)
buttons. The value can be adjusted with stepping 1mA. Press CH1 and
CH2 (▼) buttons to decrease the intensity.MIO-IONOTENS recognize
the electrodes connection: in case of faulty connection, when the
intensity reaches 15mA the value is resetted to zero.
4. The remaining time is showed on the display of MIO-IONOTENS. An
acoustic signal advises the user when the treatment is completed.
5. Turn off the device keeping pressed the
seconds.
FREE MEMORIES (ADJUSTABLE PROGRAMS)
With MIO-IONOTENS you can set the parameters according to your needs or
indicated by the doctor/physical therapist using the MEM programs.
Read the following instructions to adjust the parameters:
1. Select MEM by pressing MODE/ESC button. Scroll the programs using
PRG+ and PRG- buttons to display the preadjusted technical
specifications. Read the following instructions to adjust the chosen
program parameters: time, frequency and width impulse.
IACER Srl
button for at least two
59
button.
MNPG151-07