Automated Cleaning And Disinfecting - B.A. International BA ULTIMATE Instrucciones De Uso

Ocultar thumbs Ver también para BA ULTIMATE:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 119
6556083.book Seite 27 Montag, 18. April 2016 4:35 16
Operating Instructions BA ULTIMATE
7 Conditioning
Please observe the manufacturer's instructions for using instrument
disinfectants.
7.2

Automated cleaning and disinfecting

BA 695FBA / BA 695 can also be cleaned and disinfected in suitable
Product name
cleaning and disinfection equipment.
The cleaning and disinfection equipment used must be approved by its
Cleaning and disinfection equipment
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ BA 695FBA / BA 695 is conditioned with a cleaning and disinfection
Check _ product name
device.
1. Check whether BA 695FBA / BA 695 is clean after processing.
2. If it is still dirty, repeat the process.
27
65 56 083 D3560.201.10.02.02

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ba 695Ba 695fba

Tabla de contenido