OPERATING INSTRUCTIONS (CONT)
The 2210KL scale comes standard with these special features.
1. Auto-Zero: The scale will automatically display "0" when turned on. If the display reads
between "-20 gram" and "20 gram" at any time, regardless of what is on the tray, the scale will
automatically reset to "0" after 8 seconds.
2. English/Metric Readout: With normal operation you may switch the display units between
grams (or kilograms) and pounds at any time before or after a weighing. When turned on, the
scale will display the last used unit of measure.
3. Tare Option: An object weighing more than 20 grams can be subtracted from the total weight
to determine the net weight.
4. Weightlock: Any weight over 512 grams will lock in so that no further weighings can take
place until Weightlock is deactivated. The measurement will display even after the baby is
removed from the tray so that the caregiver can attend to the baby rather than trying to
remember the weight measurement. The value is retained even after the scale turns itself off.
5. Motion Detection: In most cases the scale will display a true, accurate weight based on a
series of consistent and stable measurements. However, if a baby or object is moving on the
platform, the scale may not achieve consistent readings that are required to give a "stable"
weight. This condition is indicated by Weightlock not activating on the display. If the movement
on the platform continues for seven seconds, the scale will automatically activate the motion
detection feature and "AVERAGE WT" will be displayed on the console.
Because of the movement on the platform, average weight indicates that the scale was not able
to achieve a true measurement in line with the listed accuracy specification. However, the
average weight deviation from the true weight of the baby will be minimal, based on the severity
of the motion during the weight reading.
6. Auto-Off: As a battery saving feature, the scale will automatically turn off when not being
used. The length of the Auto-Off timeout is programmable by the user. The factory setting is one
minute.
9