H. AUTOMATED MODE (M2, M2SU, ONE USE-PLUS, EPI-K™)
Warning: Do not connect more than one keratome at a time to the EVOLUTION 3E console.
Steps
What to do
IMPORTANT WARNINGS
1
Select the automated
Press « Mode » button (A13) until green LED (A14) is ON
mode
according the selected microkeratome.
« OP » appears on the display.
2
Proceed a vacuum test
Refer to chapter VIII.C
3
Plug the motor system
To plug One Use-Plus handpieces: A19
into the control unit
To plug M2-M2SU handpiece: A21
To plug Epi-K
4
Select the advance
The advance speed 1 is slower than the advance speed 2. The
speed (for M2, M2SU
blade oscillation rate is not modified.
and One Use-Plus)
See applicable nomograms:
• M2-M2SU : #65023
• One Use-Plus with plastic ring : #65041
• One Use-Plus with metal ring: #65067, 65077EN
• One Use-Plus for artificial chamber : #65101
5
Set the vacuum ring
Refer to user manual:
on the eye and press
• M2-M2SU: #65013
the vacuum footpedal.
• One Use-Plus with plastic ring: #65039
Assemble the keratome
• One Use-Plus with metal ring: #65040
on the ring and activate
• One Use-Plus for artificial chamber: #65009
forward footpedal to
• Epi-K™ with metal ring: #65043
perform the cut.
• Epi-K™ with plastic ring: #65044
After the cut is
Always check intraocular pressure with a tonometer (#19042) prior
performed, press
to using any keratome.
backward footpedal.
6
After surgery, switch off
the vacuum by pressing
the vacuum footpedal
and remove the vacuum
ring from the eye.
Disconnect the tubing
from the ring and the
disposable head/
blade. Discard all in an
appropriate container.
7
After use, switch off
the control unit and
disconnect the electrical
motor.
Decontaminate the
motor (see instructions in
each user manual).
handpiece: A22
TM
IX. USE OF FOOTPEDAL
A. COMPATIBILITY FOOTPEDAL-CONSOLE
Related
picture
•
Footpedals (#19361 and 19381) with serial number inferior to 7000 are not compatible with
A13
EVOLUTION 3E consoles (#19380) having a serial number superior to 7000.
A14
•
EVOLUTION 3E consoles (#19380) with serial number superior to 7000 are compatible only with
A15
footpedals (#19361 and 19381) having a serial number superior to 7000.
A8
A19
Footpedal
A21
A22
(#19361 and 19381)
A20
5000 < sn
5000 < sn < 7000
sn > 7000
B. COMPATIBILITY FOOTPEDAL-HANDPIECE
•
Footpedal #19361 can be used with M2 and One Use-Plus handpieces, as well with turbines for
ONE and CB. Footpedal #19361 can not be used with the Epi-K™ handpiece.
•
Footpedal #19381 can be used with M2, One Use-Plus and Epi-K™ handpieces, as well with
turbines for ONE and CB.
C. MANUAL MODE
C1
What to do
Press the vacuum footswitch (C1)
once: vacuum is ON
Press again the vacuum footswitch
(C1): vacuum is OFF
Press the forward footswitch (C2):
blade oscillation
B9
Press the backward footswitch (C3):
no action
EVOLUTION 3E console (#19380)
5000 < sn < 7000
compatible
compatible
not compatible
IMPORTANT WARNINGS
The system displays the actual vacuum value in mmHg (A7).
When a suction ring is fixated on the eye, the vacuum value
displayed must be less than 250 mmHg.
If the value displayed is above 250 mmHg, please refer to chapter
X.
In case vacuum is not activated, the forward footswitch does not
work.
Blade oscillation stops once forward footswitch is not pressed any
longer.
Do not press footpedal (C2) and (C3) at the same time.
#65060/INTL-J-09.2017
sn > 7000
not compatible
not compatible
compatible
Related
picture
C1
C2
C3
8