VIII. INSTALLATION AND CONNECTION
A. INSTALLATION RECOMMENDATIONS
Following is a list of general recommendations when installing the EVOLUTION 3E console:
•
Make sure the console is installed no more than 2 meters away from user
•
The console must remain in front of user during operation to visualize every indication (value,
warning)
•
User must be able to hear signals from EVOLUTION 3E console properly
•
User must be able to disconnect suction from aspiration tubing connector
•
User must be cautious about EVOLUTION 3E console displacement due to its height, dimensions
and connections
•
The EVOLUTION 3E console must not be installed above or close to a radiating device.
•
All applied parts must be used in normal condition of use, and on a routine surgical basis with
respect to their intended use according to the instructions for use.
WARNING : The microkeratomes electric motors and turbine can reach a temperature over 41°C.
Moria recommends to use the motors and rings not longer than 1 min and to prevent any contact of
motors with patient tissues.
B. ELECTRICAL SETTING
Steps
What to do
IMPORTANT WARNINGS
1
Using the screw
Finger tight only
connector, plug in the
footpedal connector
(#19361, 19381) to the
rear panel of the console
2
Connect the power
NEVER OPERATE WITHOUT MAIN POWER SUPPLY.
supply cord to the main
Before using the EVOLUTION 3E, check if position of operating
power supply
voltage (115 or 230V) is appropriate to your country's power supply:
- for a power supply of 100-120V: select position «115V»
- for a power supply of 220-240V: select position «230V».
To select position of operating voltage, see chapter XI.A.
To change fuses, see chapter XI.C.
If the unit has not been operated for four weeks or more, the battery
level should be checked and recharged prior to use.
To avoid electrical shock, the console must only be connected to a
grounded power system.
Position the console so that displays, connections and buttons
remain visible and accessible.
Always allow an easy access to the electric supply and power
socket.
In cases of emergency, disconnect the power socket or the supply
cord connector.
3
Switch on the control
Visually check all LEDs are blinking (A3, A4, A5, A7, A9, A10, A11,
unit turning B9 button
A12, A14, A15, A16, A17, A18).
to "ON"
Visually check A9 and A11 LEDs are ON.
The display indicates the atmospheric pressure (approximately
760mmHg at sea level).
C. BATTERY TEST
Steps
What to do
1
Switch EVOLUTION3E
console off by turning
button (B9) on «OFF»
Unplug the supply cord
Switch EVOLUTION3E
console ON by turning
button (B9) on «ON»
2
Switch EVOLUTION 3E
console by turning
button (B9) on «OFF»
3
Plug again the supply
cord to the unit
D. CONNECTING THE ASPIRATION TUBING
Steps
What to do
Related
1
Check the integrity of
picture
the aspiration tubing
(#19138)
B8
C
2
Connect the sterile,
single use aspiration
tubing (#19138) to the
unit:
B11
• place the chamber
in position (H1)
• then get it down
(H2)
B9
LEDS (A9
and A11)
ON
IMPORTANT WARNINGS
Check that battery charge level indicator (green LED) is switched
on (wait 15 seconds)
Check that residual gas drain switch is working (audible sound)
Perform surgery on main supply only.
IMPORTANT WARNINGS
Visually check tubing integrity before and after set up.
MORIA strongly advises against the re-use of aspiration tubing as
it may damage connector, resulting in vacuum loss or damage the
vacuum pumps.
Change tubing for each patient.
Poorly connected tubing, either on the console or on the ring, may
also cause insufficient aspiration. It is very important to check the
integrity of all connections.
Warning:
• do not reverse the chamber position of the aspiration tubing
(H3)
• if liquid appears in the chamber, replace the chamber
immediately (H4)
• visually check tubing integrity before and after set up (H5)
If moisture seeps into the console, activate vacuum to remove it.
#65060/INTL-J-09.2017
Related
picture
B9
A11
B9
B4
Related
picture
A24
H
5
MORIA