XIII. DRAWINGS
A. CONTROL UNIT (FRONT)
Figure
Description
1
Cover
2
Chamber holder
3
Pump indicator – green LED
4
Pump indicator – red LED
5
Low Vacuum LED indicator
6
Low Vacuum switch
7
Vacuum Level Indicator
8
"TEST" switch
9
Battery Level Indicators : Acceptable – green LED
10 Battery Level Indicators: Low – red LED
11 Main Power Supply Indicator – green LED
12 Manual mode indicator
13 Mode selection switch
14 Mode Display : automated
15 Mode Display and pressure display
16 Low pressure display – yellow LED
17 Correct pressure display – green LED
18 High pressure display – yellow LED
19 One Use-Plus blue connector
20 Speed 1/Speed 2 Switch (for M2, M2SU and One Use-
Plus)
21 M2 grey connector
22 Epi-K™ yellow connector
23 Turbine connector
24 Aspiration tubing connector
B. CONTROL UNIT (BACK)
Figure
Description
1
Control knob to adjust gas pressure to the turbine
2
Quick Connector for Nitrogen/Air supply
3
Connector release
4
Residual Gas Drain Switch
5
MORIA reference and Serial Number
6
Volume of Epi-K™ Footpedal Steps
7
CE mark, manufacturer address
8
Footpedal Connector
9
Main OFF (0) /ON (1) Switch
10 Fuse Housing and 115V – 230V Selector
11 Main Power Supply Connector
12 Power supply informations
13 Pumps air release
14 Screws
C. FOOTPEDAL
Figure
Description
1
Vacuum footpedal
2
Forward keratome pass or Turbine ON
3
Backward pass
4
IPX6 (serial number above 8000)
D. CONNECTING AND DISCONNECTING NITROGEN/
AIR HOSE (#19120)
Figure
Description
1
To connect : approach the hose...
2
... then screw it
3a To disconnect : first push the blue ring toward the
unit...
3b ... then pull the hose towards the back
E. HOW TO SELECT THE OPERATING VOLTAGE (115V
OR 230V POSITION)
Figure
Description
1
Switch OFF the control unit.
2
Gently open the door.
3-4 Remove the selector and select the right voltage
5
Reposition the selector and close the door
F. CHANGING THE BATTERY
Figure
Description
1
Unscrew completely the 2 screws located at the back
of the control unit
2
Slide the cover towards the front
3
Lift the cover from back to front in 2 steps (1 – 2)
4
Unplug the battery
5
Unscrew the 2 screws
6
Remove the metal piece (while keeping screws in
place)
7
Lift the battery with the plastic strap
8
Replace with a new battery, reposition and screw the
metal piece, plug the new battery, replace and screw
the cover
9
Discard the battery in an appropriate collection bin.
The used battery is classified as waste, which must
be dealt with according to the specific regulations for
collection, treatment, recycling and disposal set out in
directive 2006/66/EC of 9 September 2006 relating to
batteries and accumulators and waste batteries and
accumulators. As such the battery must be disposed of
at an appropriate collection point.
G. CHANGING FUSES
Figure
Description
1
Switch OFF the control unit.
2
Gently open the door with screwdriver
3
Remove the cartridge
4
Change the 2 fuses
5
Replace the 2 fuses
H. INSTALLING TUBING
Figure
Description
1
Place the chamber in position
2
Get the chamber down in the chamber maintener
3
Do not reverse chamber position of the aspiration
tubing
4
If liquid appears in the chamber, replace it immediately
5
Visually check tubing integrity before and after set up
11