A. HOW TO SELECT THE OPERATING VOLTAGE (115V OR 230V POSITION)
Steps
What to do
IMPORTANT WARNINGS
1
Switch OFF the control
Before opening the control unit, switch OFF the control unit and
unit.
remove the cord.
2
Gently open the door.
3
Remove the selector
For a power supply of 100-120V: select position «115V».
(A) and select the right
For a power supply of 220-240V: select position «230V».
voltage
4
Reposition the selector
and close the door
B. INSTALLING A NEW BATTERY
If the battery can not be charged, or does not hold a charge, a new battery must be installed in the console.
See draws F.
Steps
What to do
IMPORTANT WARNINGS
1
Switch OFF the control
Before opening the control unit, switch off the control unit and
unit.
remove the cord.
2
Completely unscrew the
2 screws located at the
back of the control unit
3
Slide the cover toward
the front
4
Lift the cover from back
to front in 2 steps
5
Unplug the battery
6
Unscrew the 2 screws
7
Remove the metal piece
(while keeping screws
in place)
8
Lift the battery with the
plastic strap
9
Replace with a new
Battery: 12 V - 7 Ah (Pb)
battery, reposition and
Use only the original manufacturer's parts. The use of a different
screw in the metal piece,
battery may result in severe damage to the unit, or in malfunction.
plug in the new battery,
To recycle the battery, please contact your dealer or MORIA.
replace and screw down
NEVER USE THE CONTROL UNIT WITHOUT A BATTERY
the cover
INSTALLED INSIDE AND PLUGGED IN.
Important: the used battery is classified as waste, which must
be dealt with according to the specific regulations for collection,
treatment, recycling and disposal set out in directive 2006/66/EC
of 9 September 2006 relating to batteries and accumulators and
waste batteries and accumulators. As such the battery must be
disposed of at an appropriate collection point.
10
Test battery
Refer to chapter VIII.C
C. REPLACING FUSES
Related
Steps
What to do
picture
B9
1
Switch OFF the control
unit.
B10
2
Gently open the door
E1-E2-E3
with screwdriver
E4
3
Remove the cartridge
and change the 2 fuses
4
Close the door
E5
XII. WARRANTY
A. SCOPE OF WARRANTY
DESCRIPTION
EVOLUTION 3E Console (Except Battery)
Related
EVOLUTION 3-3E Control Footpedal
picture
EVOLUTION 3E Control Footpedal Epi-K™
B9
The above items as well as spare parts and labor necessary for their repair are covered by
F1 – B14
warranty. Any items returned must be sent in their original packaging, after having previously been
decontaminated.
The maintenance operations and the replacement of spare parts will be exclusively carried out by
F2
technicians authorized by MORIA.
F3
B. NON-APPLICATION OF WARRANTY
The warranty will not be applicable under any of the following conditions:
F4
Defects or malfunction that occur out of the warranty period (paragraph XII.C.).
F5
Normal wear and tear.
Negligence or usage that does not comply with the specifications in the user manual.
F6
The use of supplies, spare parts, or accessories other than those supplied by MORIA (for example:
blades and tubing not supplied by MORIA).
Disassembly, modification, or intervention carried out on the devices by a person not authorized
F7
by MORIA.
F8
C. WARRANTY PERIOD
The warranty takes effect on the date the material is dispatched.
The duration of the warranty is 12 months from the date of effect.
D. LIABILITY
The liability of MORIA is limited to the supply of the services mentioned in paragraph XII.A. MORIA
will not be held responsible for any direct or indirect damage notably financial, suffered by the
client owing to the interventions within the scope of this warranty.
For any dispute concerning the interpretation or the execution of the present contract or the
present general terms and conditions, the Commercial Court of Nanterre, France will have sole
jurisdiction.
IMPORTANT WARNINGS
Before opening the control unit, switch OFF the control unit and
remove the cord.
500 mA high switching power
MORIA reference
19380
19361
19381
#65060/INTL-J-09.2017
Related
picture
B9
B10
G1-G2
G3-G4-G5
G1
10
MORIA

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Evolution 3e

Tabla de contenido