Flow Control; Temperature - Astralpool MAXI UP Manual De Regulatión

Ocultar thumbs Ver también para MAXI UP:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 39

FLOW CONTROL

The closing (or opening, this dipends on the programming in Utility menu, default is Normally Open
contact) of the flow input contact, free of tension, will cause:
1) the pump stops dosing
2) the green ON led blinks
3) The display shows "Flow" on the second line which alternates with the writing of the operating mode
and the flow rate.
When the flow contact returns in normal position, the pump exits from the Flow Alarm state and start
the working mode compatible with the actual inputs. If a Delay on has been set, at the end of the Flow
Alarm the pump will go on Delay on.
NOTE: The flow alarm resets the counting of the Alarm Time.
PROGRAMMING OF THE FLOW SENSOR

TEMPERATURE

PT100:
MANUAL:
Rev 0.3
In the Utility Menu, select "Flow" by the cursor and press
ENTER.
Use ▼ ▲ buttons to select the type of functioning of the flow
sensor. Press ENTER.
N.O= Normally Open (default value)
N.C= Normally Close
Press the ESC button for 3 seconds to exit programming.
In the Utility Menu, select "Temperat." By the cursor and press
ENTER.
Use ▼ ▲ buttons to select the temperature operating mode:
➔ PT100 = select PT100 if a PT100 temperature probe
is connected to input D on page 25.
Press ENTER.
➔ MANUAL = select manual if the temperature sensor
is not available. After pressing ENTER, the display
asks at what temperature is the analysis water. If you
know, set it otherwise set 25 ° C.
Press ENTER.
If you want to esc programming and return to measure, press
the ESC button for 3 seconds
Cod. 00.009.157
31

Publicidad

Tabla de contenido
loading

Tabla de contenido