TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on
Batteries have
when the START/STOP
run down.
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on display
incorrectly.
or the blood pressure
value is displayed exces-
Did you talk or move
sively low
during measurement?
(or high).
Shaking of the arm with
the cuff on.
E1 error symbol shown
Air circuit abnormality.
on display
Cuff tube may not be
plugged into monitor
correctly.
Er error symbol shown
Did you move during
on display - measure-
measurement?
ment result exceeds
Shaking of the arm with
measurement range.
the cuff on.
E2 error symbol shown
Inflation pressure exceed-
on display
ing 300 mmHg.
E3 error symbol shown
Data error.
on display
--
appears on display
Recalibration alert.
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
CORRECTION
Replace them with
four new AA alkaline
batteries.
Re-insert the batteries in
the correct positions.
Wrap the cuff properly
so that it is positioned
correctly.
Measure again. Keep
arm steady during
measurement.
Check cuff connection.
Measure again.
Measure again. Do
not move during
measurement.
Switch the unit off, then
measure again.
Remove and reload the
batteries.
Refer to page 25 for more
information.
29

SPECIFICATIONS

Four 1.5V DC (AA) batteries or
P P o o w w e e r r S S o o u u r r c c e e : :
optional 12V DC AC adapter (not included)
M M e e a a s s u u r r e e m m e e n n t t M M e e t t h h o o d d : :
Oscillometric
Pressure: 40~250 millimeters mercury (mmHg)
M M e e a a s s u u r r e e m m e e n n t t R R a a n n g g e e : :
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
A A c c c c u u r r a a c c y y : :
Pulse: ±5% of reading
P P r r e e s s s s u u r r e e S S e e n n s s o o r r : :
Semi-conductor
I I n n f f l l a a t t i i o o n n : :
Pump driven
D D e e f f l l a a t t i i o o n n : :
Automatic Pressure release valve
M M e e m m o o r r y y C C a a p p a a c c i i t t y y : :
50
memories for each User
A A u u t t o o - - s s h h u u t t - - o o f f f f : :
1 minute after last button operation
Temperature: 50°F~104°F (10°C ~40°C)
O O p p e e r r a a t t i i o o n n E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 40~85% RH max.
Temperature: 14°F~140°F (-10°C~60°C)
S S t t o o r r a a g g e e E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 10~90% RH max.
W W e e i i g g h h t t : :
0.96 lbs (without batteries)
A A r r m m C C i i r r c c u u m m f f e e r r e e n n c c e e
: :
8.7"- 18.1" (22- 46cm)
6.10"(L) x 5.12"(W) x 3"(H)
D D i i m m e e n n s s i i o o n n s s : :
155mm(L) x 130mm(W) x 74mm(H)
Type BF
Device and cuff are designed to provide special
protection against electrical shocks.
Note: These specifications are subject to change without notice.
(100
total)
30