1. Safety instructions / 2. Testing the display

1. Safety instructions

The appropriate national safety regulations (i.e. VDE 0100 /
!
EN 837-) must be observed when installing, commissioning and
GB
operating these instruments.
Caution
Only appropriately qualified personnel should work on these instruments
Pressure gauges may only be removed once they are free from pressure
Any residual pressure medium contained in the pressure element may be
hazardous or toxic. This should be taken into account when handling and
storing pressure gauges which have been removed.
Serious injuries and/or damage can occur should the appropriate regulations
not be observed

2. Testing the display

Basic testing principles
The accuracy class is tested in accordance with EN 837
The display is always tested in accordance with the operational error limit
The test is carried out with rising and falling pressure. With gauges that have
a free zero point the lower limit and the upper limit of the scale range are
used as the measuring points.
The pressure gauge used as a measuring standard must be at least four
times as accurate as the device to be tested:
e.g.: Device under test = Class 1.6
The measured values indicated are recorded after gently knocking on the
case
For pressure gauges with electrical alarm contacts, the set point indicator(s)
should be set so that any load on the actual value pointer is eliminated as
far as possible. The test of the scale range when the alarm contacts are
change-over contacts is carried out with the contacts at 1 o'clock (both
contact arms load the actual value pointer).
4
Test variable ≤ Class 0.5
WIKA Testing Instructions for Pressure Gauges