8 Maintenance
42
Does the chuck tester slide out of the chuck before the marking ring
appears?
CAUTION! The chuck is defective and the secure fit of the bur
instrument is not guaranteed. This may cause injury!
1. Do not use the product.
Workshop _ BA
2. Have the clamping system replaced by a workshop authorized by
B.A. International.
Tip: Record the time and result of the check for your own information.
8.2
Replacing the bulb in the handpiece
For turbines with W&H or Bien-Air connection, the bulb is located in the
Attachment _ W&H and BienAir
turbine handpiece.
Highspeed handpiece with W&H connection:
Highspeed handpiece_ neutral
CAUTION! The bulb may be hot. This may cause burns! Allow the
W&H T2 turbine
bulb to cool down.