Operation and function

2. Solar-thermometer
Electronic thermostat
The appliance has an electronic thermostat
that controls the temperature in the vaccine
compartment. The thermostat is set by the
factory.
1. The green diode is on when the
appliance is connected to the power
supply.
2. Solar Thermometer.
Starting procedure
Connect the mains lead to the power
supply. The green diode must light (fig. 4) to
indicate that the appliance is operating. The
electronic thermostat makes a self test, 20
seconds before the compressor starts.
Temperature control
It is recommended to check the inside
temperature with an accurate thermometer
twice a day. The inside temperature must
be checked regularly according to WHO's
standards and specifications.
4
+
°C
Cool down of the appliance
Before the appliance is loaded with
vaccines the ice-lining must be frozen. To
ensure that the ice-lining is frozen do the
following:
1. Place a thermometer in the top basket.
2. Let the refrigerator run for at least 48
Hours before loading vaccine.
3. Check the temperature in the vaccine
compartment
(must be between +2° and +8°C).
The temperature in the vaccine
compartment must always be monitored
on the thermometer and be within the
range +2° to +8°C. Due to tolerance of the
thermostat you should always control the
temperature during cool down.
8
1. Green diode
The green diode
indicates that
there is power.
fig. 8