DC/AC Current
Warning
Do not measure current when voltages might exceed 250V to avoid
blowing the fuses which can lead to,personal injury or damage to the
meter.
Current ranges are: 400.0μA, 4000.0μA, 40.00mA, 400.0mA, 10.00A.
To measure current:
1. Turn off power to circuit and discharge all capacitors.
2. Turn the rotary switch to the μA, mA or A position.
3. Press
to switch between DC/AC modes.
4. Connect the red test lead to the μAmA input jack if measured current is
less than 400mA, the A input jack for measurements above 400mA and
the black test lead to the COM jack.
5. Break the circuit and connect the leads in series across the break with the
red lead at the higher voltage side and the blck lead on the lower voltage
side.
6. Turn on power to the circuit and read the measured current on the display.
If the display shows OL the measured value has exceed the selected
range. Move the rotary switch to the next higher current position.
Specifications
General specifications
Safety Rating:
Operating altitude:
Operating Temperature/
Humidity:
Storage Temperature/
Humidity:
Temperature coefficient:
Max. Input between terminals
and earth ground:
www.grupotemper.com
CAT III, 600V; pollution degree II
<2000m
0°C to 40°, <80% RH
-10°C to 60°C, <70% RH, remove battery
0.1 x accuracy/ºC(>18ºC or <28ºC)
600V DC or AC rms
KMD 04
Digital multimeter
Instructions manual |