Indicators, Warnings And Alarms - Invacare XPO2 Manual Del Operador

Ocultar thumbs Ver también para XPO2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 106
SECTION 6—OPERATING INSTRUCTIONS

Indicators, Warnings and Alarms

Start‐Up Indicator – The XPO
 will turn On all lights briefly 
2
and check for stuck buttons upon power up. It will then 
output a brief audible beep when first turned On to indicate 
the unit is operational. 
Breath Detect Indicator ‐ The XPO
 will put out a measured 
2
pulse of oxygen each time it detects an inhalation. The blue 
flow setting indicator light will blink each time an inhalation 
is detected and the XPO
 outputs a pulse of oxygen.
2
Internal Battery Charging Indicator – When the unit is not 
turned On, but the AC, or DC, charger is connected to the unit 
and providing power, the Battery Gauge Display will 
sequentially illuminate each of the four segments. Periodically 
the sequential illumination will be interrupted by the actual 
battery charge display. If the unit is fully charged, all four 
segments will be illuminated continuously until the unit is 
turned On, or removed from charger power. If no Battery 
Gauge Display segments are illuminated, the unit is not 
charging. The lack of charging could be due to either a loss of 
power, loose charger connections or the internal battery is not 
at the proper temperature for charging. Allow the unit to 
warm, or cool, to recommended charging temperature range 
(see Recommendations for Optimal Performance on page 18), 
check for loose connections and/or try a different power outlet.
Low Battery Warning – When the XPO
 internal battery (or 
2
supplemental battery, if applicable) capacity drops to 25%, a 
brief audible beep will sound. The green 25% battery gauge 
indicator will intermittently blink.
Low Battery Alarm – When the XPO
 internal battery (or 
2
supplemental battery, if applicable) capacity drops to 15%, a 
double audible beep will sound. The green 25% battery gauge 
indicator will rapidly blink.
Part No 1160887
33
Invacare XPO2™

Publicidad

Tabla de contenido
loading

Tabla de contenido