Defender 6000 Indicators

2.2. Internal Connections

Some connections require the housing to be opened.

2.2.1 Opening the Housing

CAUTION: ELECTRICAL SHOCK HAZARD. REMOVE ALL POWER CONNECTIONS TO THE
INDICATOR BEFORE SERVICING OR MAKING INTERNAL CONNECTIONS. THE HOUSING
SHOULD ONLY BE OPENED BY AUTHORIZED AND QUALIFIED PERSONNEL, SUCH AS AN
ELECTRICAL TECHNICIAN.
i-DT61PW
1.
Remove the six Phillips head screws from the rear housing.
2.
Remove the front housing. Be careful not to disturb the internal connections.
3.
Once all connections are made, re-attach the front housing.
i-DT61XWE
1.
Remove the four hex head screws from the rear housing.
2.
Open the housing by carefully pulling the rear cover backward.
3.
Once all connections are made, re-attach the rear cover.
Note: The screws should be tightened to 2.5 N•m (20-25 in-lb) torque to ensure a watertight seal.